[Home] [Affiliations] [Diagnostics] [Facilities] [Gallery] [Personnel] [Publications] [Research] [Thrusters]

 

Time of Flight Mass Spectrometer (TOF): Determines species identification and fraction of multiply charged ions

A Time of Flight Mass Analyzer is a device often used at the end of a Parallel-Plate Energy Analyzer to distinguish between species and ionization states.  An electrostatic gate is used to close the entrance to the parallel-plate analyzer to plasma flow.  The gate is then opened momentarily (for times on the order of a few microseconds) to allow a puff of ions to travel down the drift tube.  Because ions of higher charge travel faster than singly charged ions, they will arrive at different times.  This allows the different ion charge states to be examined.

Examples of Michigan diagnostics that use a Parallel-Plate Energy Analyzer coupled with a TOF are the MBMS and ESA.

 

 


Copyright © 2006, University of Michigan
Contact the PEPL Director or the Webmaster.
Last Revised: 11/30/2006 11:56 AM